Description 

This shortwave model Crosslinker CL3000 is a microprocessor controlled UV irradiation system used among other things for UV disinfection of N-95 masks and for elimination of PCR contaminations. It combines the latest UV technology with high quality manufacturing such as an aluminum UV exposure chamber, a protective quartz disk on the UV sensor cell, and a highly resistant and easy-to-clean keypad and housing.

The CL3000 system can irradiate six masks in less than one minute. This includies the time it takes to flip the mask over, since the UVC light source is above the sample. And, because the footprint of this instrument is comparable to that of a microwave, you can easily scale up.

The CL-3000 is capable of producing a cumulative dosage of 1J/cm2, this is considered well beyond the requirement needed for disinfection. It provides near uniform UVC light with a ~10% coefficient of variation and less than 2 mJ/cm2 difference between the highest and lowest reading. Most importantly, the CL-3000 has real-time monitoring of UVC dosage to assure reproducible dosing irrespective of space and time.

Features...

  • Suitable for surface disinfection of respirator masks.
  • Precise UV dose control to deliver the required >1 J/cm2 germicidal UV dose.
  • Safety door with UV protection glass.
  • Reliable disinfection through microprocessor-controlled UV dose.

Specifications...

  • Wavelength: Short-Range Wavelength: 254 nm
  • Number of Tubes: 6 x 8 W Fluorescent bulbs
  • Instrument Dimensions (H x W x D): 8.7" x 15.7" x 13.7" (22.2cm x 40cm x 34.9cm)
  • Inside Dimensions (H x W x D): 5" x12.2" x 10.1" (12.7cm x 30.5cm x 25.4cm)
  • Weight: 5lbs (6.8kg)
  • Environmental Conditions: 15°C – 35°C, 70% air humidity, max 2000 m NN
  • UV Light:  6 x 8W 254 nm or 365 nm
  • UV Irradiation Energy: 0.1 mJ/cm to 9999.9 mJ/cm
  • Maximum Time of Exposure: 999:00 minutes

To operate, simply follow the instructions below...

  • Set the dosage on the instrument by selecting ENERGY, then pressing 6-3-0 for 63.0 mJ/cm, and then press ENTER.
  • Place mask into chamber and close door.
  • Hit START.
  • Open door and flip mask over and repeat step 3.
  • Remove the sample and continue following the UNMC protocol.
  • Note: As an added precaution, users may consider repeating steps 1-3 with an empty crosslinker in between sets of masks if residual contamination is a concern.